Programming of the indicator is divided into modular steps. Each module is a
short sequence of data entries. The front panel buttons “UP” and “DOWN”
(shown as “arrows” on the front panel) are used to change the data and set-ups,
while the “P” button is used to save or enter the data. After pressing “P”,
which gains entry into the programming mode, the programming modules are
identified by the message “Pro” and a number in the display. “UP” and
“DOWN” are used to select the desired programming module and “P” is used
to enter it. All of the subsequent programming steps follow the same procedure.
The rear terminal labeled “PGM. DIS.” must be ungrounded to gain access to
programming. The following table lists the programming steps.
“Pro 0” - RETURN TO MEASUREMENT MODE
“Pro 1” - PROGRAM THERMISTOR TYPE, TEMPERATURE SCALE
AND DECIMAL POINT POSITON
“tYPE” - Select thermistor type (400/700)
“SCALE” - Enter either degree (F or C)
“dECPNt” - Enter resolution (0 or 0.0)
“Pro 2” - PROGRAM TEMPERATURE SLOPE AND OFFSET
“SLOPE” - Enter display slope (0.0001 to 9.9999)
“OFFSEt” - Enter offset (-999 to 9999)
“Pro 3” - PROGRAM FUNCTIONS ACCESSIBLE WITH FRONT
PANEL LOCKOUT ENGAGED
“dSP AL” - Enable display alarms
“ENt AL” - Enable enter alarms =
“dSPHYS” - Enable display hysteresis
“ENtHYS” - Enable enter hysteresis =
“rSt AL” - Enable reset latched alarms
“dSPbUF” - Enable display of peak/valley readings
“rStbUF” - Enable reset of peak/valley readings=
“SELdSP” - Enable switching display between input and total
“rSttOt” - Enable reset total
“dSPOFF” - Enable display offset value
“ENtOFF” - Enable enter offset value =
“Pro 4” - PROGRAM DIGITAL FILTERING AND REMOTE
INPUT FUNCTION
“FILter” - Enter level of digital filtering
0 - no digital filtering
1 - normal filtering
2 - increased filtering
3 - maximum filtering
“E1-CON” - Enter function of remote input
0 - offset input to zero
1 - reset total
2 - reset and gate totalizer
3 - gate totalizer
4 - display hold
5 - reset peak/valley
6 - reset peak and start peak indicator
7 - reset valley and start valley indicator
8 - reset latched alarms
9 - reset all alarms
10 - toggle display between input and total
11 - offset input to zero and totalizer the offset values
12 - display hold with tare
13 - instrument reading with synchronization
14 - print request
“E2-CON” - same functions as E1-CON
APPENDIX “D” - PROGRAMMABLE FUNCTIONS